Technical Data

Measuring range 

LB 6701L-H10: 

10 µSv/h – 10 Sv/h (0°C to +50°C)

1 µSv/h – 10 Sv/h (+10°C to +40°C)

LB 6701H-H10:

100 mSv/h – 100 Sv/h (0°C to +50°C)

10 mSv/h – 100 Sv/h (+10°C to +40°C)

Energy range 45 keV – 1,3 MeV with regard to Cs-137 and 0° 
Calibration factor LB 6701L-H10: 14 fA/μSv/h
LB 6701H-H10: 1.4 fA/μSv/h 
High Voltage 1000 Volt 
Output pulse  I-F-Converter  
Polaritypositive
Amplitude+ 5 V in 50 Ω 
Pulse width 3 - 5 µs 
Frequency range100 kHz 
Operating Conditions  
Temperature0 to + 50°C 
 Rel. humidity20 to 80 % 
Storage temp.max. 60°C 
Dimensions 
Ionisation chamber80 mm Ø x 268 mm 

I-F-Converter connection box

217 x 81 x 76 mm 
Protection Degree IP 65