Highlights

  • One detector type for all: hands, feet and clothing
  • Highly robust device with transport wheels
  • Precise performance and rugged design for any environment
  • Suitable for α and β/γ contamination measurements
  • Easy to decontaminate
  • ZnS(Ag) plastic scintillation detectors
  • Very high sensitivity
  • Detachable clothing probe
  • Optical and acoustic alarm
  • Extensive nuclide library
  • Permanent data memory for >2000 entries

Product Overview

Software overview

Simple user interface and core functions

Main menu:

The clearly structured main menu allows all relevant settings to be adjusted quickly and intuitively. These include, among other functions, configuration of the measurement time, defining alarm thresholds, and adjusting the volume of the voice output.

In addition, the system provides convenient access to the integrated database. Measurement data - such as system test results - can be easily and securely exported to a USB drive as needed.

Measurement in progress:

As soon as the user enters the HFC Scan, the measurement starts automatically using the individually preset measurement time. A built-in timer on the display shows the progress in real time, ensuring full transparency. In addition, clear voice prompts guide the user safely and conveniently through the entire measurement process.

Last contamination:

If contamination is detected, the monitor provides maximum transparency: The results are clearly displayed in the memory overview, with the affected detector clearly highlighted. At the same time, the measurement value can be read quickly and easily - ensuring a safe and efficient assessment.

Technical Data

General Technical Data

Dimensions875 mm x 733 mm x 1251 mm (L x W x H, standard model)
Weightapprox. 60 kg
Housingpowder-coated steel housing
Power consumptionapprox. 120 VA

Environmental conditions

NotesNot suitable for outdoor use.
Do not operate above 2000m above sea level.
Temperature (operation)+5 °C to +45 °C
Humidity (operation/storage)75% annual average;
95% for 5 hours, non-condensing

HMI

CPUARM Cortex –A9i.MX6
Internal memory1 GB DDR3L
Operating memory4 GB eMMC
Operating systemEmbedded Linux
Interfaces1 x 12 VDC in
1 x microSD Card slot
1 x USB 2.0 OTG micro-AB
1 x USB 2.0 Type A
2 x RS-232
1 x CAN-Bus
1 x LAN
1 x Speaker out
Interfaces (right side)1 x USB 2.0
Interfaces on the back1 x LAN
Potential-free contacts:
  • 1 x Malfunction
  • 1 x Contamination
  • 1 x Heavy Contamination
  • 1 x No Contamination

Contact resistance: max. 1 A / 30 VDC

The contacts are equipped with wire- break detection (47 kΩ).

Display

Screen: 10.1” LED

Resolution: 1280 x 800 Pixel

Touchscreen: capacitive

Signal towerSignal light (optional):
• green
• yellow
• red
• Signal transmitter
• Contact voltage: 24 VDC, max. 250 mA

Detectors

Number4 for hands, 4 for feet, the left-hand or right-hand detector serves as a clothing probe
TypeSZ345, thin-layer plastic scintillation detector with ZnS coating
Active area345 cm²
Transparent protective grid83 %

Subject changes without prior notice.