Highlights

General

  • Different detector sizes and types available
  • Light, handy and extremely rugged device
  • High sensitivity and uniform response
  • Calibration factors for over 50 nuclides
  • Calibration optionally according to ISO 7503-1 or with reference to an activity of 100 cm2

SCINT Technology

  • Unique simultaneous and separate measurement of alpha and beta-gamma radiation
  • The individual types of radiation can be distinguished by means of special evaluation and correlation circuits, and thus separated and measured at the same time
  • High sensitivity even at low energies
  • No counting gas required

Probes & Connectivity

  • Additional Dose Rate Detector (LB 124 SCINT-D)
  • Xenon filled proportional counter technology (LB 124 B)
  • RS232 interface
  • Headphone connection

Different versions

LB 124 SCINT

  • Sensitive area of 170 cm²
  • Simultaneous and separate α- and β-γ-measurement
  • Calibration factors for more than 50 nuclides
  • Protective grid with high transmission
  • Adjustable alarm thresholds
  • Acoustic alarm
  • High sensitivity and uniform response
  • Lightweight, easy to handle and rugged instrument
  • Wide temperature range

Portable Contamination Monitor

The LB 124 SCINT in handheld view

Low weight, easy to handle

The bottom-side view of LB 124

Bottom side of LB 124 SCINT

LB 124 SCINT-300

 

  • Large detector area of 345 cm²
  • Time efficient measurement of large surfaces
  • High efficiency over the entire sensitive area

 

 

LB 124 SCINT-300

zoomed detector of LB 124-300

Detector area of about 345cm²

LB 124 SCINT-D

  • Mobile determination of contamination and equivalent dose rate H*(10)
  • Sensitive area of 170 cm²
  • Additional gamma dose rate detection
  • Simultaneous, separate measurement of alpha and beta-gamma radiation or gamma dose rate (2 channels)

The bottom view of LB 124 Detector

LB 124 SCINT- D with additional protective grid

LB 124-B Xenon

  • Sensitive area of 150 cm²
  • Contamination monitor on the proven detector technology with Xenon filled proportional counters
  • Precise contamination monitor, versatile and flexible in practical radiation protection
  • Offers a high sensitivity for beta radiation and photons and is particularly suitable for the detection of X-rays

Read More

 

Contamination Monitor LB 124 in laboratory

PRODUCT COMPARISON TABLE

 

LB 124 SCINT

LB 124 SCINT-300

LB 124 SCINT-D

LB 124-B

ZnS detector

X

Xenon detector

XXX

Additional dose rate detector

XXX

Weight (in g)

1300175014001620

Mechanical data
(L x W x D in mm)

240 x 140 x 110260 x 178 x 150240 x 140 x 110240 x 140 x 110

Protection class (according to IEC 60529)

IP 53IP 53IP 53IP 64

Sensitive detector area in cm²

170345170150
Radiation detectorZnS(Ag) scintillator
Measurement modes

Simultaneous alpha- and beta-gamma-measurement;
ratemeter, scaler-timer-mode, clearance measurement, survey mode

Dimensions entrance window118 mm x 145 mm
Sensitive area170 cm²
Material entrance window2 x 3 µm metallized plastic (0.4 mg/cm²)
Protective grid80 % transmission
Typical background 
(α-channel)approx. 0.1 cps
(β-γ-channel)approx. 10 cps
Typical efficiencies (according
to ISO 7503-1)

 

241Amapprox. 44 %
14Capprox. 29 %
36Cl approx. 69 %
60Coapprox. 58 %
137Csapprox. 71 %
Gamma sensitivity at 1 µSv/h 137Cs 
(α-channel)Not detectable
(β-γ-channel)< 100 cps
Spillover 
α- in β-γ-channel< 20 %
β-γ- in α-channel< 2 x 10-5
Measuring range 
(α-channel)0 to 5000 cps
(β-γ-channel)0 to 50000 cps
Radiation detectorZnS(Ag) scintillator
Measurement modes

Simultaneous alpha- and beta-gamma-measurement;
ratemeter, scaler-timer-mode, clearance measurement, survey mode

Dimensions entrance window150 mm × 230 mm
Sensitive area345 cm2
Material entrance window2 x 3 µm metallized plastic (0.4 mg/cm²)
Protective grid80 % transmission
Typical background 
(α-channel)   approx. 0.1 cps
(β-γ-channel)approx. 15 cps
Typical efficiencies (according
to ISO 7503-1)

 

241Amapprox. 46 %
14Capprox. 30 %
36Cl approx. 69 %
60Coapprox. 58 %
137Csapprox. 74 %
Gamma sensitivity at 1 µSv/h 137Cs 
(α-channel)Not detectable
(β-γ-channel)< 100 cps
Spillover 
α- in β-γ-channel< 20 %
β-γ- in α-channel< 2 x 10-5
Measuring range 
(α-channel)0 to 5000 cps
(β-γ-channel)0 to 50000 cps
Radiation detectorZnS(Ag) scintillator+ZnS(Ag) scintillator + Geiger-Müller tube
Measurement modes

α- and β-γ-measurement, simultaneous and separate, additional gamma dose rate H*(10)

ratemeter, scaler-timer-mode, clearance measurement, survey mode

Dimensions entrance window118 mm x 145 mm
Sensitive area170 cm²
Material entrance window2 x 3 µm metallized plastic (0.4 mg/cm²)
Protective grid80 % transmission
Gamma Radiation Detector 
Dose rate range 0.1 µSv/h to 20 mSv/h 
Energy range50 keV to 1.3 MeV 
Calibration factor 0.652 µSv/h per cps 137Cs
Intr. backgroundapprox. 0.07 cps 
Typical background 
(α-channel)            approx. 0.1 cps
(β-γ-channel)approx. 10 cps
Typical efficiencies (according
to ISO 7503-1)

 

241Amapprox. 44 %
14Capprox. 29 %
36Cl approx. 69 %
60Coapprox. 58 %
137Csapprox. 71 %
Gamma sensitivity at 1 µSv/h  
(α-channel)Not detectable
(β-γ-channel)< 100 cps
Spillover 
α- in β-γ-channel< 20 %
β-γ- in α-channel< 2 x 10-5
Measuring range 
(α-channel)0 to 5000 cps
(β-γ-channel)0 to 50000 cps
Radiation detectorsealed proportional counter tube with Xenon filling
Measurement modes

ratemeter, scaler-timer-mode,
clearance measurement, survey mode, half-life measurement

Dimensions entrance window100 mm x 150 mm
Sensitive area150 cm²
Material entrance window10 µm Titanium (approx. 5 mg/cm²)
Protective grid80 % transmission
Typical backgroundapprox. 10 cps
Typical efficiencies (based on the activity of a source with 100 cm² area)

 

241Amapprox. 11 %
14Capprox. 2 %
36Cl approx. 24 %
60Coapprox. 26 %
137Csapprox. 20 %

ACCESSORIES

Order Information and Accessories 

Ident. No.

Contamination monitorLB 124 SCINT43727-10
Test source

200 Bq 90Sr

800 Bq 241Am

41872

46611

Aluminum case

LB 124-KB

38164
Power supplyMultirange58067
Wall bracketfor LB 124 SCINT38789
Additional protection gridfor LB 124 SCINT45355
Data cableD-Sub cable 3 m26204
Rechargeable batteries3 x NiMH 1.2V/4.5 AH40650
Floor trolleyfor LB 124 SCINT41028
Sample holder with drawerfor LB 124 SCINT40927
Order Information and Accessories 

Ident. No.

Contamination monitorLB 124 SCINT-30048002
Test source

200 Bq 90Sr

800 Bq 241Am

41872

46611

Aluminum case

LB 124/300-KB

49700
Power supplyMultirange58067
Wall bracket

for LB 124 SCINT - 300

51374
Additional protection grid

for LB 124 SCINT - 300

49048
Data cableD-Sub cable 3 m26204

Rechargeable batteries

3 x NiMH 1.2V/4.5 AH40650
Floor trolley

for 1 x  LB 124 SCINT - 300

for 2 x  LB 124 SCINT - 300

for 3 x  LB 124 SCINT - 300

49826

56356

56946

Order Information and Accessories Ident. No.
Contamination montitor with DR detectorLB 124 SCINT-D 60026
Test source

200 Bq 90Sr

800 Bq 241Am

41872

46611

Aluminum caseLB 124-KB38164
Power supplyMultirange58067
Wall bracketfor LB 124 SCINT38789
Additional protection gridfor LB 124 SCINT45355
Data cableD-Sub cable 3 m26204
Rechargeable batteries3 x NiMH 1.2V/4.5 AH40650
Floor trolleyfor LB 124 SCINT41028
Sample holder with drawerfor LB 124 SCINT40927
Order Information and Accessories Ident. No.
Contamination monitorLB 124 B36516-10
Test source200 Bq 90Sr41872
Aluminum caseLB 124-KB38164
Power supplyMultirange58067
Wall bracket

for LB 124 B

38789
Additional protection grid

for LB 124 B

45355
Data cableD-Sub cable 3 m26204
Rechargeable batteries3 x NiMH 1.2V/4.5 AH40650
Floor trolleyfor LB 124 B41028
Sample holder with drawerfor LB 124 B40927

Applications

Nuclear Medicine
Civil Defense and Disaster Control
Research Facilities
Industry and Services
Public Facilities
Nuclear Facilities

Downloads

2 Results