Busting miths about reagent injectors
Learn more about injector qualities that have a direct impact on the performance of the assay.
PDF | 289.8 KB
Download nowWe often get asked the question what to consider when talking about the ″ideal injector”. So, let's first take a look at what we would consider to be important when taking a closer look at injectors and then compare different types of injectors against those parameters:
As you will see in the following, all injectors are not created equal. Take a look at Table 1 for a detailed comparison. Starting with a high-level view, two different designs are used, based on either syringe or bellows. What these injectors have in common is that they use a stepper motor and a 3-way solenoid valve to direct the flow of the liquid.
Parameter | Syringe-based Injector | JET Injector |
---|---|---|
Injection accuracy & reproducibility | Very High. But potential accuracy issues when injecting small volumes. Accuracy for 10µl can be as low as just 90% | Very high. 98% over the entire volume range. |
Time required per injection | Many syringe type injectors cannot perform ultra-fast injections with adequate mixing to capture the first 150ms of a kinetic assay. | Ultra-fast injection enabling the measurement of the first 150ms of a kinetic assay. |
Reagent consumption | High. High dead volume. Need to be primed up to 5 times their volume to achieve a 99% homogeneous mix at the tip. | Low. Low dead volume. Recovers up to 60% of the reagent in the injector line. A single priming sequence is sufficient to achieve a 99% homogeneous mix at the tip. |
Maintenance costs | High. Syringe based injectors are very expensive (typically >$3,000 each). Tend to wear out over time due to friction and need to be replaced. | Low. Frictionless operation for extended lifetime. Typically, more than 3 million injections without mechanical failure. |
Design & additional functions | Material mix that can produce stronger shear forces. | 100% Teflon resulting in negligible shear forces enabling injection of cell suspensions. |
The syringe-based injector is by far the most common type of injector being used today. This type of injector contains a syringe and a Teflon coated barrel. The piston moves within the glass syringe which can cause friction, resulting in increased maintenance costs.
On the other hand, most microplate readers from Berthold Technologies use up to 3 software-controlled JET Injectors which, as detailed above, have many advantages that are relevant in many types of assays. A sophisticated Prime Mode reduces the reagent consumption while ensuring homogeneous filling. JET Injectors are built for extreme longevity and fast injection times to guarantee very efficient mixing.
Learn more about injector qualities that have a direct impact on the performance of the assay.